Detail PERSONAL AIR SAMPLING MODEL : 224-PCXR8KDB (SKC INC)
Operating Range: 1000 to 5000 ml/min
(5 to 500 ml/min requires adjustable low fl ow holder)
5 to 3000 ml/min (MSHA-approved model)
Weight: 34 oz (964 gm)
Dimensions: 5.1 x 4.7 x 1.9 in (13 x 11.9 x 4.9 cm)
Compensation Range: 1000 to 2500 ml/min at 40 inches water back pressure
3000 ml/min at 35 inches water back pressure
4000 ml/min at 20 inches water back pressure
5000 ml/min at 10 inches water back pressure
Flow Control: Constant fl ow to ± 5% of set point
Run Time: 8 hours minimum at 4000 ml/min and 20 inches water back
pressure; dependent on media used
Flow Indicator: Built-in rotameter with 250 ml division;
scale marked at 1, 2, 3, 4, and 5 L/min
Battery Assembly: Plug-in battery pack, rechargeable NiCad
2.0 Ah, 6.0 V, UL Listed
Intrinsic Safety: for: Class I, Division 1 and 2, Groups A, B, C, D;
Class II, Division 1 and 2, Groups E, F, G; and Class III,
Temperature Code T3C; MSHA and ATEX-approved
models available. Contact SKC.
Operating Temp: -4 to 113 F (-20 to 45 C)
Storage Temp: -40 to 113 F (-40 to 45 C)
Charging Temp: 41 to 113 F (5 to 45 C)
Humidity: 0 to 95% Relative
Flow and Fault shutdown at excessive back pressure or low battery
Low Battery Fault: with LCD indicator and time display retention
Battery Test: LCD Indicator verifi es battery condition prior to sampling.
Time Display: LCD Indicator displays time in minutes for sampling period
elapsed time, pump run time, or total elapsed time including
delayed start time.
Delay On: Allows user to select minutes to delay start of test up to
9999 minutes (7 days)
Timed Shutdown: Allows user to select minutes of operation before
automatic shutdown
Timing Accuracy: ± 0.05% (± 45 sec/day)
Intermittent Sampling: User-programmable to extend short-term samples over
an extended period of time to meet time-weighted average
(TWA) requirements with a reduced number of samples.
Elapsed time maximum is 9999 minutes (7 days).
Sampling Pause: Allows user to temporarily halt sampling without loss
of timing or programming data; restart does not require
resetting time
Multiple-tube Sampling: Built-in constant pressure regulator allows user to take
up to four simultaneous samples with fl ow rates of up to
500 ml/min each (total combined flow 1350 ml/min
maximum) using optional adjustable low fl ow holder.
RFI Shielding • Complies with requirements of EN 55022, FCC Part 15
Performance: Class B, EN 50082-1
• Frequency range of the radiated susceptibility test was
27 MHz to 1000 MHz
• marked
Additional Specs: • Built-in fl uid/particulate trap
• External see-through fi lter housing
• Tamper-resistant control cover
• Tamper-resistant compensation adjustments
• Built-in accessory mount
• Stainless steel belt clip
• External exhaust port
• All non-plastic parts made of corrosion-resistant materials
• RFI and EMI-shielded case